Vanessa Tan Yee Jueen


Session

09-10
08:45
75min
Imaging In CI
Laila Telmesani, Yutaka TAKUMI, Mahmoud Mandour, Badr Eldin Mostafa, Nozomu Matsumoto, Vanessa Tan Yee Jueen, Levent Sennaroğlu, Luminita Radulescu

Cochlear implants (CIs) were initially incompatible with MRI. Recently, MRI-conditional implants that permit scanning without magnet removal or compressive bandaging have been introduced. However, MRI in CI recipients is often degraded by device-related artifacts, which may limit its clinical utility. In this study, we evaluated head MRI findings in some patients who underwent cochlear implantation at our institution and analyzed differences in artifact size across imaging modalities and sequences, together with their clinical applicability. Without magnet removal, diffusion-weighted imaging (DWI) and T2 star-weighted imaging produced larger artifacts and yielded limited diagnostic value. In contrast, T1-weighted, T2-weighted, fluid-attenuated inversion recovery (FLAIR), and heavily T2-weighted sequences allowed assessment of the contralateral side and central regions of the head, although visualization on the implanted side remained restricted. Image characteristics varied depending on both MRI modality and sequence, indicating that MRI protocols should be carefully selected according to clinical purpose and feasibility. Awareness of sequence-specific imaging characteristics is essential, and the expected diagnostic value should be considered in advance when determining MRI indications in CI patients.

Cochlear Implants and implantable devices
Hearing Implant 3 (ICC - B3 Floor - 3B/66)